Medical device supplier audit checklist
Quality management system is the regulatory requirement for a medical device manufacturer, and it is foundation of the quality of a medical device manufacture. What’s the key points we need to check in the medical device manufacture audit?
Quality Organization | |
Is there a quality manual defining the company’s policies, objectives, and commitment to quality? (please attach copy) | |
Does the manual describe the Quality Department and its relationship to the rest of the organization? | |
Does the manual identify specific persons by title for the responsibility of each quality function? | |
Have adequate resources, including trained personnel, been identified and assigned for management and performance of work and verification activities? | |
Is there a Management Representative appointed by the company’s executive management with defined and written authority for the implementation and performance evaluation of the quality system? | |
Are Management Reviews conducted at defined intervals to review the quality system to ensure the continuing suitability and effectiveness of the quality system? | |
Quality System | |
Are quality system procedures documented and maintained to ensure product conformance with specified requirements? | |
Is the Quality Management System continually monitored and updated to remain current with industry practices and regulatory expectations? | |
Are quality plans developed to ensure that activities associated with the design, material procurement, fabrication, inspection and release of product meet the requirements specified in the contract/purchase order? | |
Does Quality Assurance participate in contract review to identify and evaluate quality requirements? | |
Internal Audits | |
Are there planned and executed internal audits to determine the effectiveness of the quality system? | |
Are internal audits conducted by appropriately trained personnel who are independent of the activities and functions being audited? | |
Are internal audit results reported to responsible personnel? | |
Are corrective actions assigned accordingly and verified for effectiveness? | |
Training | |
Do qualifications for jobs affecting quality include identification of appropriate education, training needs, and experience? | |
Is a documented training program in place to ensure adequate training for personnel in relation to their job function(s)? | |
Are training records maintained? | |
Purchasing Controls | |
Are purchasing procedures documented and maintained to ensure purchased product conforms to specified requirements? | |
Is there a documented procedure for the qualification, use, and monitoring of approved suppliers? | |
Are suppliers required to send notification for approval of changes prior to implementation? | |
Do purchasing documents contain adequate descriptions and specifications of products being ordered? | |
Are contracts and purchase orders reviewed to ensure that customer requirements are clearly defined and achievable? | |
Are records of contract/purchase order reviews maintained? | |
Design Controls | |
Are procedures documented and maintained to ensure product conformance through the control and verification of design? | |
For each design and development activity, are plans prepared, documented, and maintained? | |
Are risk analysis tools employed throughout the design process? | |
Are formal, documented reviews of design results planned and conducted at appropriate stages of design? | |
Are design verifications and validations performed at appropriate stages of design to ensure that the design outputs meet the requirements of documented design inputs? | |
Are design changes and modifications identified, reviewed, and approved prior to their implementation? | |
Are all records related to the design process maintained in a design history file (DHF)? | |
Currently a supplier to CTC and using CTC Controlled Documents? Are Design Document revisions current with CTC DMR document revisions? Annual recertification of CTC Documents submitted to and approved by CTC? | |
Change Control | |
Is there a written procedure describing customer notification for review and approval of changes to material, components, the manufacturing process, equipment, location, cost and delivery? | |
Document Control | |
Is there a documented process for the creation, revision, approval, distribution, and obsolescence of procedures? | |
Are documents and data reviewed for adequacy and approved by authorized personnel prior to release? | |
Is there a document master list that identifies the current revision status of controlled documents? | |
Are procedures readily available to users, with revised and obsolete documents being promptly removed to prevent unintended use? | |
Are changes reviewed and approved by the same functions that performed the original review and approval? | |
Are procedures documented and maintained for the identification and control of quality records? | |
Are records maintained that demonstrate product conformance and effective operation of the quality system? | |
Are quality records readily retrievable and stored in secure facilities that prevent damage, deterioration, and loss? | |
Are electronic data backed up at appropriate intervals with backup media stored in secure facilities that prevent damage, deterioration, and loss? | |
Are record retention periods defined & documented? | |
Manufacturing Process Controls | |
Do controlled procedures exist for manufacturing operations, describing process controls and specifications? | |
Are line clearances performed and documented prior to manufacturing startup? | |
Are inspections of first and last articles performed and documented at manufacturing startup and end? | |
Are procedures in place for the identification and handling of raw materials, work in progress, and finished product? | |
Are workmanship criteria developed, documented, and implemented? | |
Is a bill of materials established for each product? | |
Is there a documented procedure addressing the approval, verification, and validation of any changes? | |
Are production records maintained and reviewed for compliance? | |
Are procedures in place for maintaining facility and product cleanliness? | |
Is there a documented preventive maintenance system ensuring that all maintenance activities are recorded and reviewed? | |
If any environmental controls are applied during manufacturing, are associated records maintained? | |
Process Validation | |
Are written procedures in place for the qualification and validation of equipment and processes? | |
Are written procedures in place for the validation of software, including software used as part of the quality management system?? | |
Are validation plans and specifications established with results of validation activities documented and reported? | |
Have all processes been validated? | |
Are changes to processes, equipment, and software validated prior to release? | |
Inspection & Testing | |
Are procedures documented and maintained for product inspection and testing to ensure specifications are met? | |
Are sampling plans established for inspection and testing? | |
Are purchased materials inspected or verified for conformance prior to use? | |
Is in-process testing conducted in accordance with established procedures or quality plans that are based on customer requirements? | |
Is the final product inspected and tested in accordance with established procedures or quality plans that are based on customer requirements? | |
Are the results of all inspections, testing, and any related analyses documented and maintained in device history records? | |
Is there a final review of all documentation related to production, inspection, and testing prior to final product release? | |
Are written procedures in place for the identification and segregation of product by status (i.e. Accepted, Rejected, Quarantined, etc.)? | |
Measuring & Test Equipment | |
Are written procedures established for the control, calibration, and maintenance of equipment used for inspection, measuring, and testing? | |
Is equipment calibrated at regular intervals with calibration covering the range of use? | |
Is equipment uniquely identified to show the calibration status? | |
Are calibrations traceable to the National Institute of Standards and Technology (NIST) or another nationally recognized standard? | |
Are records of calibration activities and results maintained? | |
Are procedures established for the assessment of product impact in the event that any equipment falls outside of specified limits? | |
Is calibration performed internally or by a third party? | |
Nonconforming Product | |
Are written procedures established for the identification, documentation, evaluation, segregation, and disposition of nonconforming product? | |
Is a documented procedure established for reprocessing or rework of nonconforming product? | |
Has the responsibility for review and authority for disposition of nonconforming product been defined? | |
Are records related to the investigation and disposition of nonconforming product maintained? | |
Corrective & Preventive Action | |
Have documented procedures been established for the implementation of corrective and preventive actions? | |
Are procedures established for the effective handling and documentation of customer complaints? | |
Are procedures established addressing the investigation, root cause analysis, and corrective action required for nonconformities related to product, process, and quality system? | |
Do procedures for preventive action address the use of appropriate sources of information to detect, analyze, and eliminate potential causes of nonconformities? | |
Are documented procedures in place addressing the implementation and effectiveness checks of corrective and preventive actions? | |
Are records of corrective and preventive actions documented and maintained? | |
Material Controls | |
Are written procedures established for the receipt, handling, storage, labeling, packaging, and delivery of product? | |
Are product handling, storage, and delivery methods developed to prevent damage and deterioration? | |
Are labeling and packaging operations controlled to prevent mix-up? | |
Is there an individual designated to review and proof labeling for accuracy? | |
Are storage areas clearly identified with products appropriately segregated (such as incoming raw material or finished product, nonconforming product, usable stock)? | |
Where required, is the storage environment controlled? | |
Is traceability maintained on each lot of raw material to its applicable certification or test report? | |
Are “First In / First Out” (FIFO) procedures maintained for stock control? | |
Are procedures established for the control and handling of outdated/expired materials? | |
Are unique identifiers assigned to incoming or in-process materials? | |
Are lot numbers representing one homogenous production run assigned to final products? | |
Are materials shipped with records identifying product, quantity, and conformance? | |
Statistical Techniques | |
Are statistical techniques used in establishing, controlling, and verifying process capability and product characteristics? | |
Is the use of statistical tools documented? |